Качество Karl Storz 27005BA Cystoscope for repair Фабрика

Karl Storz 27005BA Cystoscope for repair

Karl Storz 27005BA Cystoscope for repair ► Brief Description Fault: 1. No image. 2. Fiber abrasion. 3. Rod lens broken and cracked. 4. Objective lens broken and dirty. 5. Eyepiece cover broken. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens ...
Качество LINVATEC HD 4300 Arthroscope for repair Фабрика

LINVATEC HD 4300 Arthroscope for repair

LINVATEC HD 4300 Arthroscope for repair ► Brief Description Fault: Image with black shadow; eyepiece with scratches, inner optical parts broken. Solution: Replace objective lens, inner optical parts, clean and adjust optical system. ►Problem & Solutoin Brand LINVATEC Model HD 4300 Fault Image with ...
Качество Stryker 30º 0502-503-030 laproscope for repair Фабрика

Stryker 30º 0502-503-030 laproscope for repair

Stryker 30º 0502-503-030 laproscope for repair ► Brief Description Fault: Image blur, objective lens loose without glue, inner optical parts broken. Solution: Add glue for objective lens, replace inner optical parts, clean and adjust optical system, seal with eyepiece. ►Problem & Solutoin Brand ...
Качество Olympus A70940A Автоклавируемый артроскоп для ремонта Фабрика

Olympus A70940A Автоклавируемый артроскоп для ремонта

Olympus A70940A Autoclavable Arthroscope for repair ► Brief Description Fault: 1. Image blur, black dots and dark. 2. Fiber not light enough. 3. Rod lens broken. 4. Objective lens broken and dirty. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens ...
Качество Гастроскоп Olympus GIF-Q150 для ремонта Фабрика

Гастроскоп Olympus GIF-Q150 для ремонта

Olympus GIF-Q150 Gastroscope for repair ► Brief Description Fault: 1. With image; 2. Light bundlebreak badly; 3. Insert tubeaging; 4. Biopsy tubeabrasion, leakage badly; 5. Angle tied, bendingsection, wires and helix tube abrasion; 6. Rubberwithout glue. Solution: Replace light bundle, insert tube, ...
Качество Гастроскоп GIF-XQ240 производства Olympus для ремонта Фабрика

Гастроскоп GIF-XQ240 производства Olympus для ремонта

GIF-XQ240 Gastroscope by Olympus for repair ► Brief Description Fault: Leakage. Solution: Replace light guide tube,NO.1 button,wire,mesh,adjust angle. ►Problem & Solutoin Brand Olympus Model GIF-XQ240 Fault Leakage. Solution Replace light guide tube,NO.1 button,wire,mesh,adjust angle. ► Common ...
Качество Цифровой цистоскоп Olympus CYF V2 для ремонта Фабрика

Цифровой цистоскоп Olympus CYF V2 для ремонта

Olympus CYF V2 Cystoscope For Repair ► Brief Description Fault: Leakage. Solution: Replace the insert tube, light bundle, biopsy channel, bending section, wire, steel mesh, rubber. ►Problem & Solutoin Brand Olympus Model CYF V2 Fault Leakage Solution Replace the insert tube, light bundle, biopsy ...
Качество Гастроскоп Olympus GIF-H170 для ремонта Фабрика

Гастроскоп Olympus GIF-H170 для ремонта

Olympus GIF-H170 gastroscope for repair ► Brief Description Fault: 1. The top left and middle right side with blur image, objective Lens broken. 2. 20% light bundle broken, the small lens turned yellow. 3. Insert tube yellow and wrinkles. 4. The biopsy tube abrasion. 5. The wire and helix tube ...
Качество Р Вольф 7326071 уретероскоп для ремонта Фабрика

Р Вольф 7326071 уретероскоп для ремонта

R Wolf 7326071 ureteroscope for repair ► Brief Description Fault: The image has spots,image bundle aging,light bundle broken;Insert tube aging;Biopsy channel worn and leakage;Bending section,wire and helix tube were worn. Solution: Replace the light bundle,image bundle,rubber,biopsy channel,helix ...
Качество Ларингоскоп Insighters IS3-F для ремонта Фабрика

Ларингоскоп Insighters IS3-F для ремонта

Insighters IS3-F laryngoscope for repair ► Brief Description Fault: 1. Image blank screen; 2. Insert tube abrasion 3. Bending section transformation; 4. Distal head transformation; Solution: Replace CCD, insert tube, bending section, rubber ►Problem & Solutoin Brand Insighters Model IS3-F Fault 1. ...
Качество КАРЛ СТОРЗ 11278А1 уретероскоп для ремонта Фабрика

КАРЛ СТОРЗ 11278А1 уретероскоп для ремонта

KARL STORZ 11278A1 ureteroscope for repair ► Brief Description Fault: 1) Image spotted with partially black point, image bundle aged 2) Light bundle partially wire broken 3) Insert tube aged 4) Biopsy tube abrasion and leakage 5) Wires and coil pipe abrasion Solution: Replace HD image bundle, light ...
Качество Фиброназофарингеальный ларингоскоп Olympus ENF-GP для ремонта Фабрика

Фиброназофарингеальный ларингоскоп Olympus ENF-GP для ремонта

Olympus ENF-GP fiber nasopharyngeal laryngoscope for repair ► Brief Description Fault: 1. Image with black point and image bundle broken 2. Light bundle aged 3. Insert tube aging and wrinkle 4. Bending section deformation 5. Angle with not enough 6. Rubber abrasion Solution: Replace image bundle; ...
Качество Бронхоскоп Olympus BF-XP160F для ремонта Фабрика

Бронхоскоп Olympus BF-XP160F для ремонта

Olympus BF-XP160F fiber scope for repair ► Brief Description Fault: 1. Image degumming, serious color, CCD aging, a few black and gray spots; 2. 20% of the light bundle is broken; 3. Insert tube aging slightly; 4. Biopsy tube wear, not smooth; 5. Bending section, steel wire, screw tube wear; ...
Качество Олимпус CF-2T160L Колоноскоп для ремонта Фабрика

Олимпус CF-2T160L Колоноскоп для ремонта

Olympus CF-2T160L Colonoscope for repair ► Brief Description Fault: 1. With image. 2. 10% light bundle broken. 3. Insert tube is ok. 4. Biopsy tube abrasion. 5. Angle tied, bending section transformation. 6. Wires, helix tube abrasion, no up angle. Solution: Replace image bundle; insert tube; biopsy ...
Качество Олимпус CF-Q160AL Колоноскоп для ремонта Фабрика

Олимпус CF-Q160AL Колоноскоп для ремонта

Olympus CF-Q160AL scope for repair ► Brief Description Fault: 1.No image, CCD broken; 2.image broken completely; 3.insert tube aging; 4.biopsy tube abrasion; 5.bending section abrasion; 6.wires and helix tube abrasion; 7.light guide tube aging and wrinkle. Solution: Replace CCD, light bundle, insert ...
Качество Эндоскоп Olympus CHF-P20Q для ремонта Фабрика

Эндоскоп Olympus CHF-P20Q для ремонта

Olympus CHF-P20Q endoscope for repair ► Brief Description Fault: 1.Insert tube cracked; 2.Rubber leakage; 3.Image with black point and cracked; 4.Angulation tight; 5.Biopsy tube aging Solution: Replace image bundle; insert tube; biopsy tube; bending mesh and wires; coil pipe; rubber ►Problem & ...
Качество Гастроскоп OLYMPUS GIF-1TQ160 для ремонта Фабрика

Гастроскоп OLYMPUS GIF-1TQ160 для ремонта

Olympus GIF-1TQ160 gastroscope for repair ► Brief Description Fault: 1.With image, but CCD with broken. 2.Light bundle broken completely. 3.Insert tube abrasion. 4. No biopsy tube. 5. No air&water tube, no water supply tube. 6. No bending section assembly. 7. No digital tip. Solution: Replace CCD, ...
Качество Гастроскоп Olympus GIF-H180J для ремонта Фабрика

Гастроскоп Olympus GIF-H180J для ремонта

Olympus GIF-H180J gastroscope for repair ► Brief Description Fault: 1.image with shadow; 2.light bundle broken; 3.insert tube with bite mark; 4.biopsy tube abrasion; 5.bending section serious deformation; 6.wires and helix tube abrasion; 7.light guide tube aging and wrinkle. Solution: Replace CCD, ...
Качество Гастроскоп OLYMPUS GIF-Q180 для ремонта Фабрика

Гастроскоп OLYMPUS GIF-Q180 для ремонта

Olympus GIF-Q180 gastroscope for repair ► Brief Description Fault: 1.image definition not enough, CCD aging. 2.30% light bundle broken. 3. Insert tube aging and wrinkle. 4.Biopsy tube abrasion. 5. Bending section bited, transformation. 6. Angle not enough, wires and helix tube abrasion. 7. Light ...
Качество Олимпус CF-240AI Колоноскоп для ремонта Фабрика

Олимпус CF-240AI Колоноскоп для ремонта

Olympus CF-240AI Colonoscope for repair ► Brief Description Fault: Image blur,insert tube aging,light guide squashed,light guide broken 50%,nozzle blocked,angle tight,forceps channel aging. Solution: Replace insertion tube,CCD chips,light guide,forceps channel,water channel,angulation wire,bending ...
Качество Торакоскоп Olympus LTF-160 для ремонта Фабрика

Торакоскоп Olympus LTF-160 для ремонта

Olympus LTF-160 thoracoscope for repair ► Brief Description Fault: 1.with image, but big risk for CCD after leakage. 2. Light bundle broken completely. 3. Insert tube with little scratch. 4. Biopsy tube abrasion and leakage. 5. Wires and helix tube abrasion. 6. Grip broken, leakage. 7. All the ...
Качество Эндоскоп Olympus TJF-160VR для ремонта Фабрика

Эндоскоп Olympus TJF-160VR для ремонта

Olympus TJF-160VR endoscope for repair ► Brief Description Fault: 1.No image, CCD broken. 2. Light bundle broken completely. 3. Insert tube aging and wrinkle. 4. Biopsy tube broken. 5. Wires and helix tube broken. 6. clamp wires broken. 7. Operation part broken, no angle. 8. Grip broken. 9. The ...
Качество Гастроскоп Olympus GIF-Q150 для ремонта Фабрика

Гастроскоп Olympus GIF-Q150 для ремонта

Olympus GIF-Q150 gastroscope for repair ► Brief Description Fault: 1.with image; 2.one eye of light bundle broken; 3.insert tube little aging; 4.biopsy tube abrasion; 5.bending section deformation; 6.wires and helix tube abrasion; 7.light guide tube aging and wrinkle. Solution: Replace light bundle ...
Качество STORZ 13820pks Гастроскоп для ремонта Фабрика

STORZ 13820pks Гастроскоп для ремонта

STORZ 13820pks Gastroscope for Repair ► Brief Description Fault: 1.Rubber leakage; 2.Wire abrasion; 3. no up angle, other angle lack of. Solution: 1. Replace the wires, coil pipe, rubber; 2.Adjust angle. ►Problem & Solutoin Brand STORZ Model 13820pks Fault 1.Rubber leakage; 2.Wire abrasion; 3. no up ...